Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Sub Soltec TJ1 (1 year) CAL VER /5520 DATE: 11-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 20 NUMBER OF LINES: 125 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Soltec TA100-438/TA200-938 Instruction Manual, TJ1 Specifications # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- N K P F V 1.003 ASK+ R X 1.004 HEAD INITIAL SETUP 1.005 DISP On the Mainframe, set the following... 1.005 DISP 1.005 DISP TRIGGER: MANU. TRIG. 1.005 DISP Y SIZE: 200mm 1.005 DISP T SIZE: X1 1.005 DISP TIME/DIV: 2 sec. 1.006 DISP EXTERNAL CONTROL settings on the Mainframe... 1.006 DISP MODE: REAL 1.006 DISP TRIGGER: SHOT 1.007 DISP On the TJ1, set the following... 1.007 DISP 1.007 DISP LINE WIDTH: MEDIUM 1.007 DISP TRIGGER: OFF 1.007 DISP LOWER TEMP (C): -100 1.007 DISP RANGE (C): +100 1.008 HEAD {ZERO} 1.009 DISP Connect the 5520A TC terminals to the TJ1 INPUT 1.009 DISP using Type J thermocouple wire and connectors. 1.010 DISP Press START on the Mainframe to feed the chart paper. 1.011 HEAD ZERO: {-100degC} 1.012 5520 -100degC _J TC S 2W 1.013 EVAL Does TJ1 {indicate 0% deflection (-100degC)}? 2.001 HEAD ZERO: {-50degC} 2.002 DISP LOWER TEMP (C): -50 2.003 5520 -50degC _J TC S 2W 2.004 EVAL Does TJ1 {indicate 0% deflection (-50degC)}? 3.001 HEAD ZERO: {0degC} 3.002 DISP LOWER TEMP (C): 0 3.003 5520 0degC _J TC S 2W 3.004 EVAL Does TJ1 {indicate 0% deflection (0degC)}? 4.001 HEAD ZERO: {50degC} 4.002 DISP LOWER TEMP (C): 50 4.003 5520 50degC _J TC S 2W 4.004 EVAL Does TJ1 {indicate 0% deflection (50degC)}? 5.001 HEAD ZERO: {100degC} 5.002 DISP LOWER TEMP (C): 100 5.003 5520 100degC _J TC S 2W 5.004 EVAL Does TJ1 {indicate 0% deflection (100degC)}? 6.001 HEAD ZERO: {200degC} 6.002 DISP LOWER TEMP (C): 200 6.003 5520 200degC _J TC S 2W 6.004 EVAL Does TJ1 {indicate 0% deflection (200degC)}? 7.001 HEAD ZERO: {300degC} 7.002 DISP LOWER TEMP (C): 300 7.003 5520 300degC _J TC S 2W 7.004 EVAL Does TJ1 {indicate 0% deflection (300degC)}? 8.001 HEAD ZERO: {400degC} 8.002 DISP LOWER TEMP (C): 400 8.003 5520 400degC _J TC S 2W 8.004 EVAL Does TJ1 {indicate 0% deflection (400degC)}? 9.001 HEAD ZERO: {500degC} 9.002 DISP LOWER TEMP (C): 500 9.003 5520 500degC _J TC S 2W 9.004 EVAL Does TJ1 {indicate 0% deflection (500degC)}? 10.001 HEAD ZERO: {600degC} 10.002 DISP LOWER TEMP (C): 600 10.003 5520 600degC _J TC S 2W 10.004 EVAL Does TJ1 {indicate 0% deflection (600degC)}? 11.001 HEAD {FULL SCALE} 11.002 DISP On the TJ1, set the following... 11.002 DISP 11.002 DISP LOWER TEMP (C): 0 11.002 DISP RANGE (C): +100 11.003 HEAD FULL SCALE: {100degC} 11.004 MESS Slew for Full Scale (200mm) Deflection. 11.005 ASK- R 11.006 5520 100 100degC 0.3% _J TC 2W 12.001 ASK+ R 12.002 HEAD FULL SCALE: {200degC} 12.003 5520 200 200degC 0.3% _J TC 2W 13.001 HEAD FULL SCALE: {500degC} 13.002 5520 500 500degC 0.3% _J TC 2W 14.001 HEAD FULL SCALE: {1000degC} 14.002 5520 1000 1000degC 0.3% _J TC 2W 15.001 HEAD FULL SCALE: {1200degC} 15.002 5520 1200 1200degC 0.3% _J TC 2W 16.001 MESS 16.002 HEAD {LINE WIDTH} 16.003 DISP Set the LINE WIDTH control to OFF. 16.004 EVAL [N]Is {there a line recording}? 17.001 DISP Set the LINE WIDTH control to LIGHT. 17.002 EVAL Is the {line recording Light}? 18.001 DISP Set the LINE WIDTH control to MEDIUM. 18.002 EVAL Is the {line recording Medium}? 19.001 DISP Set the LINE WIDTH control to BOLD. 19.002 EVAL Is the {line recording Bold}? 20.001 DISP Press STOP on the Mainframe to stop the chart paper. 20.002 END